Digiflex Com is a compact high performance instrument designed for the location of faults on copper cables. Its capabilities are marked out particularly by the short range resolution (down to 2.5cm) and the long range of 30km.

Ease of use and the incorporation of the latest technology enable even inexperienced users to achieve quick and precise locations with the Digiflex Com. For the expert a multitude of more advanced features is available.

The proven method of pulse reflection is employed whereby a string of suitable pulses is sent down the cable which are wholly or partially reflected back from the fault. The size and shape of these indicate the type of fault.

Performance Characteristics:

  • High resolution of location down to 2.5 cm on a 50 m range
  • Ease of operation through use of menu and help functions
  • Adjustable range dependent amplitude correction
  • Clear evaluation of the trace through simultaneous display of zoom and full screen (split-screen)
  • Large area LC-Display  can be backlit
  • Single, continuous and average test modes
  • electrical strength of test inputs
  • Built in memory capacity for  50 traces
  • Serial connection to printer and PC
  • Records from PC can be uploaded to the instrument memory using Database software (Winkis)
  • Robust, shower proof  housing  for rough field use

Optional accessories:

  • Mains supply unit 
  • Transport case   
  • Fault Converter (for location of moisture faults)
  • Cable phantom KNB001
  • Replacement accumulator    
  • Delivery with calibration certificate

Technical Data

Range: 50 m to 30000 m (with zoom 5 m)
Types of test: Direct test with internal reconstitution, IFL-Mode (intermittent faults)
Comparison of two pairs
Differential display of two pairs
Coupling (cross talk) tests, conductor crosses,
Ingress of moisture with Fault Converter
Pulse widths: 5, 10, 20, 50, 200, 750, 3000 ns
Distance measurement in: µs, metres or feet
Max. resolution: 2.5 cm
Accuracy: ± 0.1 % of range
± cursor steps
V/2 settings: 30 - 150 m/µs
98 - 492 ft/ms
NVP-Factor: 0.200 - 0.9999
Dynamic range: > 90 dB
Display: 256 x 128 Pixel
Pulse amplitude: 10 V into 130 Ohm
Internal reconstitution: 40 - 200 Ohm
Power supply: NiMH-replaceable accumulator and mains,
230 V AC; 11 - 15 V DC
Op. temperature: 10 °C to +50 °C
Storage temperature: -20 °C to +60 °C
Housing: Plastic to IP 55 under DIN 40 050
Dimensions (L x H x D): 260 x 176 x 70 mm
Weight: 1.5 kg